Sample is introduced
Water is supplied to the analyser within the recommended sample flow of 0–2 L/min and pressure of 0–1 bar.
Monitor essential water quality parameters with a UV correlation-based analyser designed for wastewater and water treatment applications. Measure BOD, COD and TSS, with pH and optional flow-rate monitoring.
UV spectroscopy-based measurement supports routine monitoring of key water quality indicators. Automatic cleaning and a long-life lamp help keep routine maintenance focused on periodic replenishment of the cleaning solution.
The Metazone Online Water Analyser is primarily based on UV spectroscopy, a measurement approach known for stability and low operating cost. It is intended for applications where regular observation of water quality can help operators assess treatment performance and reduce the risk of pollutant discharge.
External probes can be added for additional physicochemical measurements such as pH, ORP, dissolved oxygen, conductivity and turbidity, according to the required configuration.
Key operating and performance specifications of the Online Water Analyser. Optional parameters and configurations can be discussed according to the application requirement.
| Parameter | Details |
|---|---|
| Parameters Supported | BOD, COD, TSS, pH, Flow Rate (Optional) |
| Additional Parameters | As per requirement |
| BOD Range | 0–500 mg/L |
| COD Range | 0–500 mg/L |
| TSS Range | 0–500 mg/L |
| pH Range | 2–12; 0–14 (Optional) |
| Measurement Method | UV Correlation |
| Parameter Accuracy | ±5% |
| pH Accuracy | ±0.1 pH |
| Recommended Sample Flow | 0–2 L/min |
| Recommended Sample Pressure | 0–1 bar |
| Operating Temperature | 0–70 °C |
| Sample Interval | Customizable, 5 minutes and above |
| Cleaning Interval | Customizable, 2 minutes and above |
| Data Logging | Internal |
| Memory | 8 GB |
| Recommended Maintenance Interval | 3 months |
| HMI Size | 7.0-inch touchscreen |
| HMI Resolution | 800 × 480 pixels |
| Communication | RS485 |
| Telemetry | Yes, via RTU |
| Certification | CE |
| Enclosure | Mild steel |
| Dimensions | 800 × 600 × 210 mm |
Specifications are based on the supplied product datasheet. Final configuration may vary with selected options and project requirements.
The analyser uses UV spectroscopy and UV correlation for water quality measurement. A sample is presented to the measurement system at the recommended flow and pressure conditions, and the instrument processes the measurement for display and logging.
Water is supplied to the analyser within the recommended sample flow of 0–2 L/min and pressure of 0–1 bar.
The UV spectroscopy-based system determines supported water quality parameters through its configured measurement process.
Measurements are available on the 7-inch touchscreen HMI, with internal data logging and RS485 communication for remote monitoring.
Configure the analyser around the water quality indicators and process conditions relevant to your site. Additional parameters and external probes can be considered as per requirement.
BOD, COD and TSS are supported in the stated 0–500 mg/L ranges. pH measurement is supported in the 2–12 range, with 0–14 available as an option.
Flow Rate can be included as an optional measurement parameter where required by the monitoring application.
External probes can be added for pH, ORP, dissolved oxygen, conductivity and turbidity, subject to the selected configuration.
Sample intervals are customizable from 5 minutes and cleaning intervals from 2 minutes and above.
Designed for water quality monitoring across municipal, natural water and industrial wastewater environments.
A water quality monitoring solution with features intended to support routine observation, data access and maintenance planning.
Monitor BOD, COD, TSS and pH through one analyser platform, with optional flow-rate measurement.
Automatic cleaning and a long-life lamp help limit routine maintenance, with periodic cleaning-solution replenishment.
Internal data logging and 8 GB memory provide onboard capacity for recorded measurement data.
Connect the analyser to a remote monitoring arrangement using the available communication interface and telemetry option.
The MODBUS interface is available over RS485. A master controller can poll the instrument to read current instantaneous readings from the available channels for remote monitoring applications.
Telemetry is supported via RTU, enabling integration into a suitable remote monitoring architecture according to the project design.
Refer to the accessory details below when planning installation and site readiness.
| Accessory | Provision / Details |
|---|---|
| Sampling Pump | Included (optional); cleaning solution is in the end-customer's scope |
| Cleaning Pump | Internal |
| Power Cord | Included |
| Remote Terminal Unit | Included |
| 8 mm Fitting Pipe | 10 m |
Metazone Engineers works with customers to understand the application and align the instrument configuration with site requirements.
Discuss your monitoring parameters, sample conditions and communication requirements with Metazone Engineers.